7-3
IM 704210-01E
Displaying the Measured/Computed Data
7
Explanation
Selecting the histogram display function
From the display format menu, select HISTOGRAM.
Performing auto scaling
The X-axis and Y-axis will be set automatically so that the entire distribution of measured data
can be displayed.
The X-axis and Y-axis scaling can be changed as described below.
Setting the X-axis and Y-axis
• Setting range for center value of X-axis.
The setting range varies according to the selected measurement function.
• Period, time interval or pulse width measurement: 10 ns to 99.9999900 ms (up to 3.19
µ
s in
the case of hardware histogram mode)
• Phase difference measurement:
15.00 to 345.00 deg
• Duty ratio measurement:
0.500 to 99.500%
• Selecting the span of X-axis
The span varies according to the selected measurement function.
• Period, time interval or pulse width measurement: 2/4/10/20/40/100/200/400 ns, 1/2/4/10/20/
40/100/200/400
µ
s, 1/2/4/10/20/40/100 ms (up to 3.2
µ
s in the case of hardware histogram
mode)
• Phase difference measurement: 30/60/90/180/360 deg
• Duty ratio measurement: 1/5/10/50/100%
• Selecting the Y-axis scale.
• LIN: Linear scale
• LOG: Logarithmic scale
• Selecting the maximum value of Y-axis
The maximum value of Y-axis varies according to the selected Y-axis scale.
• LIN: 10/20/40/100/200/400/1000/2000/4000/10000/20000/40000/1e5/1e6/1e7/1e8 (up to
32000 in the case of time stamp mode)
• LOG: 1e1/1e2/1e3/1e4/1e5/1e6/1e7/1e8 (e.g. 1e2 indicates 10
2
.) (up to 1e5 in the case of time
stamp mode)
7.1 Setting the Histogram Display Function and Reading the Data (Readout)