1-11
IM 704210-01E
Functions
1
Inhibit
≡
For operating procedure, refer to 6.4
≡
Measurement can be inhibited by inputting the inhibit signal to the INHIBIT IN terminal. This is
also possible while the gate is open or during measurement following the arming delay.
Measurement can be inhibited for up to 320 seconds. In addition, it is possible to select whether
to inhibit measurement using the positive or negative edge of the inhibit signal.
1.5 Setting the Acquisition Conditions for Input Signal to be Measured
Relation between the inhibit signal and gating
S1
S2
S3
Input signal
S4
S5
Not measured.
Gate
Inhibit signal
Mesurement is inhibited.
Gate is open.
Start of measurement
Measurement is resumed.
Not counted as an
event during inhibition
of measurement.
Relation between the inhibit signal and gating/external arming
S1
S2
S3
S4
S5
External arming
Input signal
Not measured.
Gate
Inhibit signal
Mesurement is inhibited.
Gate is open.
Start of measurement
Measurement is resumed.
Not counted as an
event during inhibition
of measurement.
Relation between the inhibit signal and gating/external arming/arming delay
S1
S2
S3
S4
S5
Arming delay
External arming
Input signal
Not measured.
Gate
Inhibit signal
Mesurement is inhibited.
Gate is open.
Start of measurement
Measurement
is resumed.
Not counted as an
event during inhibition
of measurement.