10-7
IM 704210-01E
Using Other Functions
10
10.3 Outputting the Monitor Signal
≡
≡
≡
≡
≡
For a description of the function, refer to 1.7
≡
≡
≡
≡
≡
Connecting a BNC cable
Connect a BNC cable to the monitor output terminals (MONITOR OUT CHA/CHB) provided on
the rear panel of the instrument.
Explanation
The measuring signal input to CHA is output from the MONITOR OUT CHA terminal
constantly, and the measuring signal input to CHB is output from the MONITOR OUT CHA
terminal constantly.
(CHA monitor
output terminal)
(CHB monitor
output terminal)
Monitor output circuit
CH
A
CH
B
MONITOR OUT
( 50 )
( 50 )
50
Ω
+5V
–5V
• Output impedance: 50
Ω
(typical *)
• Output voltage level: Approx. 1/4 of the measuring input signal (within
±
5 V) when the output
is received by input impedance of 50
Ω
* Typical value is described in terms of typical of average performance. However, they cannot
be guaranteed.
CAUTION
Never short-circuit the monitor output terminals or apply an external voltage to them.
Damage to the internal circuits may result.