2-48
IM DL350-03EN
2.15 Connecting a Logic Probe to the Instrument’s
Logic Input Terminal or the Logic Input Module
To measure logic signals, connect a logic probe (the 702911, 702912, 700986, or 700987) to the
instrument’s logic input terminal or a logic input module (the 720230).
CAUTION
• Applying a voltage greater than the limits listed below may damage the logic probe or the
instrument. If the frequency is above 1 kHz, damage may occur even when the voltage is
below this value.
Maximum input voltage (at a frequency of 1 kHz or less)
• Logic probes (702911 and 702912): 35 V
• High-speed logic probe 700986: 42 V (DC + ACpeak)
• Isolated logic probe 700987:
250 Vrms (however, ACpeak must be less than 350 V,
and DC must be less than 250 V)
• For logic probes (702911 and 702912), and high-speed logic probe 700986, the eight input
lines of a single pod share the same GND. Also, the instrument’s GND and the GNDs of
each pod are connected. Do not apply signals with different common voltages to each
input line. Doing so may damage the instrument, connected logic probes, and connected
devices.
• The input terminals of an isolated logic probe are isolated from each other and from the
instrument.
• Turn off the instrument before you connect or remove a 26-pin connector from the logic
signal input connector.
• Do not stack the isolated logic probes during use. Also, allow enough space around the
probes to avoid a temperature increase inside the probes.
• Do not use the YOKOGAWA 700985 logic probe with the instrument. The 700985 is
shaped so that it can be connected to the logic signal input connector of the instrument, but
it is not electrically compatible with the instrument, so connecting the two could damage the
instrument or the 700985.
French
ATTENTION
• Le fait d’appliquer une tension supérieure aux limites mentionnées ci-dessous risque
d’endommager la sonde logique ou l’instrument. Si la fréquence est supérieure à 1 kHz,
des dommages risquent de survenir même lorsque la tension est inférieure à cette valeur.
Tension d’entrée maximale (à une fréquence de 1 kHz ou moins)
• Sondes logiques 702911 et 702912 :
35 V
• Sonde logique à grande vitesse 700986 : 42 V (c.c. + crête c.a.)
• Sonde logique isolée 700987 :
250 Vrms (cependant, la valeur crête c.a. doit
être inférieure à 350 V et c.c. à 250 V)
• Pour les sondes logiques (702911 et 702912), et la sonde logique à grande vitesse
700986, les huit lignes d’entrée d’une cosse unique partagent la même terre (GND). De
plus, la GND de l’instrument et les GNDs de chaque cosse sont reliées. Ne pas appliquer
de signaux de tensions communes différentes vers chaque ligne d’entrée. Cela pourrait
endommager l’instrument, les sondes logiques reliées et les appareils reliés.