6-1
IM DL35O-03EN
Specifications
1
2
3
4
5
6
App
Index
6.1 Signal Input Section
Item
Specifications
Type
Plug-in input unit
Number of slots
2
Maximum number of input
channels
8 channels (when 4-CH modules are installed in both slots) + instrument’s standard logic 16 bit
32 channels (when 16-CH modules are installed in both slots) + instrument’s standard logic 16 bit
240 channels (when the 720240 or 720241 modules are installed in both slots) + instrument’s
standard logic 16 bit
Memory size
200 Mpoints total (100 M per slot)
6.2 Recorder Mode Feature
Waveform Acquisition and Display
Item
Specifications
Record conditions
Preset time recording
Records data for the specified time period from the start point
Continuous recording
Records data for the specified time period before stopping
Trigger start
Records data for the specified time period from a trigger
Trigger end
Records data for the specified time period before a trigger
Acquisition mode
Normal
Normal waveform acquisition
Envelope
Holds peak values at the maximum sample rate, regardless of the time
axis setting
Recording time
10 s to 50 days
Sampling interval
1 μs to 200 ms (1-2-5 series)
Action performed at the end
of recording
Screen capture data saving, waveform data saving, buzzer notification, mail transmission
SD recording (binary format) Sampling interval
Depends on the number of used channels, 10 μs minimum (when 10
channels are in use)
1
Maximum number of
recorded points
1 Gpoint (with limitations depending on the used module)
Operation description
Stores data in binary format in sync with acquisition
SD recording (ASCII format) Recording interval
1, 2, 5, 10, 15, 20, 30 s, 1, 2, 5, 10, 15, 20, 30, 60 min
Maximum recording size
2 GByte
Operation description
Stores data in text format every specified recording interval
Event recording
Records up to 100 events using the event input terminal
Display time range
1 ms to 10 s (1-2-5 steps), 20 s, 30 s, 40 s, 50 s, 60 s, 100 s, 200 s, 300 s
10 min to 60 min (10 min steps), 100 min
2 hour, 5 hour, 10 hour to 60 hour (10 hour steps), 80 hour, 100 hour
5 day, 10 day, 20 day, 30 day,
2
40day,
2
50 day
2
Zoom
One window
Display format
1, 2, 3, 4, 5, 6, 8,12, 16 divided windows of TY display
Maximum number of
displayed traces
32 (standard logic 16 bit , including Math)
X-Y display
Select X and Y axes from analog input waveforms and Math waveforms, up to two traces in one
window
1 May become 10 μs or more depending on the SD card
2 Only during real-time recording
Vertical Control
Item
Specifications
Vertical axis setting
Set with the measurement range
Channel on/off
CHn, CHn_m, and Mathn can be turned on and off separately.
Vertical axis zooming
By setting the scale using upper and lower limits
Linear scaling
Can be set to Ax+B mode or P1-P2 mode (only for voltage, stress, and frequency)
Chapter 6
Specifications