6-11
IM DL35O-03EN
Specifications
1
2
3
4
5
6
App
Index
Item
Specifications
Emissions
Compliant standards
EN 61326-1 ClassA , EN 61326-2-1, EN 55011 : Class A,Group1
EMC Regulatory Arrangement in Australia and New Zealand EN 55011 Class A,
Group 1
Korea Electromagnetic Conformity Standard (
한국 전자파적합성기준
)
This product is a Class A (for industrial environment) product. Operation of this
product in a residential area may cause radio interference.
Further, when a measurement lead or probe is connected to an input module or
when connected to the device under measurement, the emission requirements
may be exceeded.
In such cases, the user will be required to correct the interference.
Cable conditions
Functional ground cable
Use a cable that is 3 m or less in length to connect to ground.
DC power cable
Use a cable that is 3 m or less in length. Do not connect to a DC distribution
network.
Ethernet port (Ethernet)
Use a category 5 or better STP cable that is 30 m or less in length.
USB port for PCs
Use a shielded USB cable that is 3 m or less in length.
I/O port
Use a cable that is 3 m or less in length.
Immunity
Compliant standards
EN 61326-1 Table 2 (for use in industrial locations), EN 61326-2-1
When a measurement lead or probe is connected to an input module or when
connected to the device under measurement, the immunity requirements may
not be met.
In such cases, the user will be required to correct the interference.
Influence in the immunity environment
701261 ≤ ± 3 mV
701262 ≤ ± 3 mV
701265 ≤ ± 0.05 mV
701270 ≤ ± 500 μSTR
701271 ≤ ± 500 μSTR
701275 ≤ ± 6 mV
720211 ≤ ± 50 mV
720220 ≤ ± 20 mV
720221 ≤ ± 2 mV
720230 No bit error
720240 No error
720241 No error
720243 No error
720250 ≤ ± 20 mV
720254 ≤ ± 15 mV
720266 ≤ ± 0.05 mV
720268 ≤ ± 100 mV
720281 ≤ ± 0.01 Hz
Cable conditions
Same as the emission cable conditions
Environmental standard
Compliant standard
EN 50581 Monitoring and control instruments including industrial monitoring
and control instruments.
Vibration resistance standard
JIS D 1601:1995 5.2 5.3 (1) Type 1: Type A compliant
Function: 10 m/s
2
, 5-400 Hz, XYZ axes 10 minutes each
Durability: 30 m/s
2
, 33 Hz, Z axis 4 hours, X and Y axes 2 hours each
1 Battery pack cover (720923) is required for battery pack operation.
6.12 General Specifications