2-9
IM DL350-03EN
Making Preparations for Measurements
1
2
3
4
5
6
App
Index
2.4 Installing Input Modules
Types of Input Modules
The following types of input modules are available.
High-Speed 100 MS/s, 12-Bit Isolation Module
720211 (HS100M12)
High-Speed 10 MS/s, 12-Bit Isolation Module
720250 (HS10M12)
4-CH 1 MS/s, 16-Bit Isolation Module
20254 (4CH 1M16)
Hi-Voltage 1 MS/s 16bit Isolation Module (with AAF, RMS)
720268 (HV (AAF, RMS))
Universal (Voltage/Temp.) Module
701261 (UNIVERSAL)
Universal (Voltage/Temp.) Module (with AAF)
701262 (UNIVERSAL (AAF))
Temperature, High Precision Voltage Isolation Module
701265 (TEMP/HPV)
Temp./Hi-Prec. Voltage Module (Low Noise)
720266 (TEMP/HPV)
Strain Module (NDIS)
701270 (STRAIN_NDIS)
Strain Module (DSUB, Shunt-Cal)
701271 (STRAIN_DSUB)
Acceleration/Voltage Module (with AAF)
701275 (ACCL/VOLT)
Frequency Module
720281 (FREQ)
16-CH Voltage Input Module
720220 (16CH VOLT)
16-CH Temperature/Voltage Input Module
720221 (16CH TEMP/VOLT)
Logic Input Module
720230 (LOGIC)
CAN Bus Monitor Module*
720240 (CAN MONITOR)
CAN & LIN Bus Monitor Module*
720241 (CAN & LIN)
SENT Monitor Module*
720243 (SENT)
* The CAN bus monitor, CAN&LIN bus monitor, and SENT monitor modules can be used on models with the /VE
(vehicle edition) option.
Precautions to Be Taken When Installing or Removing Input
Modules
If you replace one installed input module with another, the settings for the channel will be reset to their
defaults when the power is turned on. If you want to keep the settings, specify a save destination and
save them.
Installation Procedure of Modules
1.
Make sure that the instrument is turned off.
2.
Check the channel number displayed on the input module installation slot on the top panel of
the instrument, and then install the input module along the guide.
Holding the handles on the top and bottom of the input module, press hard until it clicks in
place. If there is a cover panel on the slot in which to install the module, remove the cover
panel, first.
3.
Firmly fasten the screws that came with the instrument in two places: the top and bottom of the
input module.
Screw tightening torque: 0.6 N•m
4.
Turn the instrument’s power switch on.
5.
In the overview screen, check that the name of the module that you installed is displayed
correctly at the appropriate slot. If it is not correct, remove the module according to the steps
in “Removal” shown below, and reinstall the module according to steps 1 to 3 shown above. To
display the overview screen, see section 5.6, “Viewing System Information (Overview).”
Guide