App-14
IM DL35O-03EN
Appendix 2 Relationship between the Main Channel
Sample Rate, Sub Channel Data Update
Rate, and Acquisition Memory Writing Rate
16-CH Voltage Input Module
Rate at Which Data Is Written to the Acquisition Memory
Number of Sub Channels Used
1
2
3 or 4
5 to 8
9 to 16
Sub Channel Sample Rate
(Data Update Rate)
200 k
100 k
50 k
20 k
10 k
Main Channel
Sample Rate Setting (S/s)
100 M
500 k
500 k
500 k
500 k
500 k
50 M
500 k
500 k
500 k
500 k
500 k
20 M
500 k
500 k
500 k
500 k
500 k
10 M
500 k
500 k
500 k
500 k
500 k
5 M
500 k
500 k
500 k
500 k
200 k
2 M
500 k
500 k
500 k
200 k
100 k
1 M
500 k
500 k
200 k
100 k
50 k
500 k
500 k
100 k
100 k
50 k
20 k
200 k
200 k
100 k
50 k
20 k
10 k
100 k
100 k
50 k
20 k
10 k
5 k
50 k
50 k
10 k
10 k
5 k
2 k
20 k
20 k
10 k
5 k
2 k
1 k
10 k
10 k
5 k
2 k
1 k
500
5 k
5 k
1 k
1 k
500
200
2 k
2 k
1 k
500
200
100
1 k
1 k
500
200
100
50
500
500
100
100
50
20
200
200
100
50
20
10
100
100
50
20
10
5
50
50
10
10
5
2
20
20
10
5
2
1
10
10
5
2
1
–
5
5
1
1
–
–
: Some data will not be updated. If this occurs, previous data is displayed consecutively.
If you set the sample rate (the main channel sample rate setting) of this instrument to a rate that
exceeds the sub channel data update rate, all data within the sub channel data update interval will
be the same.