6-12
IM DL35O-03EN
6.13 Module Specifications
High-Speed 100 MS/s, 12-Bit Isolation Module (720211)
Item
Specifications
Standard operating
conditions
Temperature:
23 ± 5°C
Humidity:
20 to 80% RH
After a 30 minute warm-up and after calibration
Effective measurement
range
20 div (display range: 10 div)
Number of input channels
2
Input coupling
AC, DC, GND
Maximum sample rate
100 MS/s
Input type
Isolated unbalanced
Frequency characteristics
1
(−3 dB point a when sine
wave of amplitude ±3 div is
applied)
DC to 20 MHz
Voltage-axis sensitivity
setting
10 mV/div to 20 V/div (1-2-5 steps) (when using 1:1 probe factor)
Maximum input voltage
(at a frequency of 1 kHz or
less)
Combined with the 700929 (10:1), 702902 (10:1), or 701947 (100:1):
2
1000 V (DC+ACpeak) CAT II
Combined with the 701954 (1:1):
4
200 V (DC+ACpeak)
Direct input (cable that does not comply with the safety standards):
6
42 V (DC + ACpeak)
Maximum rated voltage to
ground
Between input and case
Between inputs
(at a frequency of 1 kHz
or less)
Working voltage of safety standard
Combined with 700929 (10:1) or 701947 (100:1),
3
or combined with 701954 (1:1:)
5
1000 Vrms (CAT II)
Combined with the 702902 (10:1):
3
1000 V (DC+ACpeak) CAT II
Direct input (cable that does not comply with the safety standards):
7
42 V (DC + ACpeak) (CAT II, 30 Vrms)
Vertical (voltage) axis
accuracy
DC accuracy
1
10 mV/div to 20 V/div: ±(0.5% of 10 div)
Input connector
BNC connector (isolated type)
Input impedance
1 MΩ ±1%, approx. 35 pF
Low frequency –3 dB point
when AC coupled
10 Hz or less (1 Hz or less when using the 700929 or 702902, 0.1 Hz or less when using the
701947)
Common mode rejection
ratio
80 dB (50/60 Hz) or more (typical value
8
)
Residual noise level (input
section shorted)
±1.1 mV or ±0.15 /div whichever is greater (typical value
8
)
Withstand voltage
5400 Vrms for 2 seconds (across each terminal and earth) (60 Hz)
Allowable transient surge
voltage (instantaneous)
±2100 Vpeak (across each input terminal and earth)
Insulation resistance
500 VDC, 10 MΩ or more (across each input terminal and earth)
A/D converter resolution
12 bit (150 LSB/div)
Temperature coefficient
Zero point:
10 mV/div to 20 V/div: ±(0.1% of 10 div)/°C (typical value
8
)
Gain:
±(0.05% of 10 div)/°C (typical value
8
)
Bandwidth limit
Select from Full and 2 MHz
Cut-off characteristics:
−12 dB/OCT (typical value
8
)
Digital filter
Cutoff frequency:
1.28 MHz, 640 kHz, 320 kHz, 160 kHz, 80 kHz, 40 kHz, 20 kHz, 10 kHz
Filter format:
IIR
Probe attenuation setting
Voltage probe: 1:1, 10:1, 100:1, 1000:1
Current probe: 1 A:1 V, 10 A:1 V (for the 701932 or 701933), 100 A: 1 V (for the 701930 or 701931)
1 Value measured under standard operating conditions