6-40
IM DL35O-03EN
CAN Bus Monitor Module (720240)
Item
Specifications
Standard operating
conditions
Ambient temperature: 23 to 5°C
Ambient humidity:
20 to 80% RH
Number of input ports
2
Maximum sample rate
100 kS/s
Input format
Isolated unbalanced
Maximum input voltage
−3 V to +10 V (between CAN_H and GND or CAN_L and GND)
Maximum rated voltage to
ground
Between input and case
Between inputs
42 V (DC + ACpeak) (CAT II, 30 Vrms)
Input connector
D-sub 9 pin (male)
Terminator
Internal. Can be enabled or disabled for each port
Supported protocols
Physical layer: ISO-11898 (High Speed Communication)
Supported bit rates
1
10 kbps, 20 kbps, 33.3 kbps, 50 kbps, 62.5 kbps, 66.7 kbps, 83.3 kbps, 100 kbps, 125 kbps,
250 kbps, 500 kbps, 800 kbps, 1Mbps
LED display
Illuminates to indicate that the terminator is enabled
Number of sub channels
60 sub channels/ports
One shot output
Frames can be output in single shots.
Terminator resistance
(when the terminator is
enabled)
110 to 130 Ω
A/D converter resolution
(when Value Type is set to Float)
16 bits, 2400 LSB/div
1 Low-speed CAN at low-speed transceiver levels is not supported.
6.13 Module Specifications