16-2
IM 253710-01E
16.2 Error Messages and Corrective Actions
Error Messages
There are cases in which error messages are displayed on the screen during operation.
This section describes the meanings of the messages and the appropriate corrective
actions. The messages can be displayed in English or Japanese (see section 15.2). If
the corrective action indicates servicing, contact your dealer for repairs.
In addition to the error messages listed in this section, there are also communication
related error messages. These messages are described in the Communication Interface
User’s Manual (IM253710-11E).
Error in Execution
Codes Messages
Actions
Sections
10
Occure ACQ time out.
—
—
11
Cannot measure PLL frequency.
Check input level.
6.4
601
Invalid file name or SCSI ID.
Check file name or SCSI ID.
12.3, 12.5
602
No media inserted or no SCSI device.
Make sure that the storage medium is
12.1 to 12.3
inserted (if applicable), and check the SCSI
device connection and the SCSI ID.
603
No SCSI device or no media inserted.
Check the SCSI device connection and the SCSI
12.1 to 12.3
ID, and make sure that the storage medium is
inserted (if applicable).
604
Media failure.
Check the storage medium.
—
605
File not found.
Check the file name and the storage medium.
—
606
Media is protected.
Set the disk’s (medium’s) write protect switch to OFF.
—
607
Media was removed while accessing.
Check the storage medium.
—
608 ,
File already exists.
—
12.5
609
610
Contains invalid characters.
—
12.5
611,
Media full.
Delete unnecessary file(s) or use another disk.
12.4 to 12.7
612
613
File system failure.
Check using another disk. If the same message
—
still appears, maintenance service is required.
614
File is protected.
—
12.8
615
Physical format error.
Reformat the medium. If the same error occurs, the
12.4
instrument is probably unable to execute a
format on this medium.
616
File system failure.
Check using another disk. If the same message
—
to 620
still appears, maintenance service is required.
621
File is damaged.
Check the file.
—
622
File system failure.
Check using another disk. If the same message
—
to 641
still appears, maintenance service is required.
642
No media exists in SCSI device.
Check that the storage medium is correctly inserted
—
in the SCSI device.
643
Media failure.
Check the medium.
—
to 653
654
Media failure.
Check the format type of the floppy disk.
—
656
File system failure.
Check using another disk. If the same message
—
to 664
still appears, maintenance service is required.
665
Cannot load this file format. File was stored
—
—
on other models or other versions.
666
File is now being accessed.
Execute after access is made.
—
667
Cannot be executed while running.
Press START/STOP key to stop acquisition.
4.3
668
Cannot find ‘.HDR’ file.
Check the file.
12.6