10-23
IM 253710-01E
Numerical Computations
10
Explanation
Select the range of harmonic orders that will be used to determine the numerical data of
the measurement functions of the harmonic waveform.
Selecting the minimum harmonic order under analysis
Select from the following choices.
• 0
Includes the 0th order (DC) component when determining the various numerical
data of the harmonic waveform.
• 1
Does not include the 0th order (DC) component when determining the various
numerical data of the harmonic waveform. 1st order (fundamental signal) is the
first component used.
Selecting the maximum harmonic order under analysis
Select the harmonic order in the range from 1 to 500th order.
Note
•
When the minimum harmonic order is set to [1], the numerical data of Total of the harmonic
measurement data will not include the DC component.
•
You can set the maximum harmonic order to be up to 500th order, but the maximum
harmonic order that is actually used to determine the numerical data is the order
corresponding to the upper limit of the harmonic order under analysis. This upper limit is
automatically determined (maximum is 500) from the frequency of the PLL source. The
numerical data corresponding harmonic orders exceeding the upper limit are shown as [-------]
(no data).
10.7 Setting the Harmonic Orders under Analysis