17-16
IM 253710-01E
Item
Specification
Current input
Input terminal type
Direct input : large binding post
Current sensor input : BNC connector
Input format
Floating input, shunt input method
Input impedance
Direct input (5 A)
Input resistance : Approx. 100 m
Ω
Input reactance : Approx. 0.07
µ
H
Current sensor input
Input resistance : Approx. 10 k
Ω
Measurement range
Direct input (5 A)
Select from Auto, 10 Apk, 4 Apk, 2 Apk, 1 Apk,
0.4 Apk, 0.2 Apk, 0.1 Apk.
Current sensor input
Select from Auto, 1000 mVpk, 400 mVpk, 200
mVpk, 100 mVpk.
Frequency range
DC to 2 MHz
Instantaneous maximum Direct input (5 A)
30 Apk or 15 Arms, whichever is less.
allowable input (1 s)
Current sensor input
2 Vrms
Continuous maximum
Direct input (5 A)
10 Apk or 7 Arms, whichever is less.
allowable input
Current sensor input
2 Vrms
Continuous maximum
600 Vrms (50/60 Hz)
common mode voltage
CMRR
Open input terminals, apply 600 Vrms (50/60 Hz) between input
(Influence from common terminals and case, f is the frequency in kHz.
mode voltage)
0.005% of range or less at 10 Hz
≤
f
≤
1 kHz.
(Maximum range / measurement range)
×
0.0002
×
f% of range or less
for other frequencies (designed value). f is the frequency in kHz.
Line filter
Select from OFF, 500 Hz, 20 kHz, and 1 MHz.
Zero crossing filter
Select from OFF, 500 Hz, and 20 kHz.
A/D converter
12 bit
Sampling rate
5 MS/s maximum
17.15 Power Measurement Module (253751)