9-9
IM 253710-01E
Waveform Display
9
Explanation
The screen can be evenly divided and the waveform corresponding to each channel can
be assigned to one of the divided windows.
Selecting the number of divisions
Select the number of divisions from the following list of choices.
• Single : No division
• Dual : Two divisions
• Triad : Three divisions
• Quad : Four divisions
Depending on the number of divisions, the number of displayed points in the vertical
direction varies as follows. The display resolution of sampled data, such as the
voltage level, does not change (see section 1.6).
Single : 432 points, Dual : 216 points, Triad : 144 points, Quad : 108 points
Waveform mapping
• Auto
The waveforms of CH1, CH2, ..., Math1, Math2 are assigned in order from the top
divided window. The channels that are turned OFF are skipped.
• Fixed
The channels that are turned OFF are also assigned to the windows. Math1 and
Math2 are displayed in the top window and 2nd window, respectively.
When the screen is divided into four windows [Quad], CH1 through CH6, CH8,
and Math2 are ON, and CH7 and Math1 are OFF.
CH1, CH5
CH2, CH6, Math2
CH3
CH4, CH8
CH1, CH5
CH2, CH6
CH3, CH8
CH4, Math2
Fixed
Auto
• User
The channels can be assigned arbitrarily to the divided windows regardless of
whether or not the display is turned ON. You can select the display position from 0 to
3. The channels are assigned in order starting from number 0 from the top divided
window. This is applicable to products (PZ4000) with firmware version 2.01 or later.
User (when the screen is divided into three windows
with CH2, Math1, and Math2 set to 0, 1, and 3, respectively)
CH2, Math2
Math1
0, 3
1
2
The channels are displayed in
order according to the specified number.
9.3 Splitting the Screen and Displaying the Waveforms