5-4
IM 253710-01E
• When the power measurement modules are installed in element number 1, 2, 3,
and 4 slots
Selectable
Wiring Method
Numerical Data of the
Σ
Function
A-(B)
1P2W-1P2W
•
Σ
A : Displays the numerical data of element 1.
•
Σ
B : Displays the numerical data of element 2.
* Elements 3 and 4 are also handled as 1P2W. They do not affect
Σ
A and
Σ
B.
1P2W-1P3W
•
Σ
A : Displays the numerical data of element 1.
•
Σ
B : Displays the numerical data when elements 2 and 3 are combined into
a 1P3W system.
* Element 4 is handled as 1P2W. It does not affect
Σ
A and
Σ
B.
1P2W-3P3W
•
Σ
A : Displays the numerical data of element 1.
•
Σ
B : Displays the numerical data when elements 2 and 3 are combined into
a 3P3W system.
* Element 4 is handled as 1P2W. It does not affect
Σ
A and
Σ
B.
1P2W-3V3A
•
Σ
A : Displays the numerical data of element 1.
•
Σ
B : Displays the numerical data when elements 2, 3, and 4 are combined
into a 3V3A system.
1P2W-3P4W
•
Σ
A : Displays the numerical data of element 1.
•
Σ
B : Displays the numerical data when elements 2, 3, and 4 are combined
into a 3P4W system.
1P3W-1P2W
•
Σ
A : Displays the numerical data when elements 1 and 2 are combined into
a 1P3W system.
•
Σ
B : Displays the numerical data of element 3.
* Element 4 is handled as 1P2W. It does not affect
Σ
A and
Σ
B.
1P3W-1P3W
•
Σ
A : Displays the numerical data when elements 1 and 2 are combined into
a 1P3W system.
•
Σ
B : Displays the numerical data when elements 3 and 4 are combined into
a 1P3W system.
1P3W-3P3W
•
Σ
A : Displays the numerical data when elements 1 and 2 are combined into
a 1P3W system.
•
Σ
B : Displays the numerical data when elements 3 and 4 are combined into
a 3P3W system.
3P3W-1P2W
•
Σ
A : Displays the numerical data when elements 1 and 2 are combined into
a 3P3W system.
•
Σ
B : Displays the numerical data of element 3.
* Element 4 is handled as 1P2W. It does not affect
Σ
A and
Σ
B.
3P3W-1P3W
•
Σ
A : Displays the numerical data when elements 1 and 2 are combined into
a 3P3W system.
•
Σ
B : Displays the numerical data when elements 3 and 4 are combined into
a 1P3W system.
3P3W-3P3W
•
Σ
A : Displays the numerical data when elements 1 and 2 are combined into
a 3P3W system.
•
Σ
B : Displays the numerical data when elements 3 and 4 are combined into
a 3P3W system.
3V3A-1P2W
•
Σ
A : Displays the numerical data when elements 1, 2, and 3 are combined
into a 3V3A system.
•
Σ
B : Displays the numerical data of element 4.
3P4W-1P2W
•
Σ
A : Displays the numerical data when elements 1, 2, and 3 are combined
into a 3P4W system.
•
Σ
B : Displays the numerical data of element 4.
5.2 Selecting the Wiring Method