App-22
IM 253710-01E
For harmonic measurement mode
• When the measurement function is U, I, or P (The measured values of the harmonics and the harmonic
distortion factors are saved.)
Address
Numerical Data
Address
Numerical Data
0000
Measured value Total
—
—
0004
Measured value DC component
07D8
Harmonic distortion factor of the DC component
0008
Fundamental component
07DC
Harmonic distortion factor of the fundamental component
(1st order)
000C
2nd order harmonic component
07E0
Harmonic distortion factor of the 2nd order harmonic component
↓
↓
↓
↓
07D4
500th order harmonic component
0FAC
Harmonic distortion factor of the 500th order harmonic component
Note
Even if the upper limit of the harmonic order under analysis is less than 500th order, data up
to 500th order are saved. The data of the harmonic orders that are not analyzed are saved
as NAN (0x7FC00000).
• When the measurement function is other than those indicated above and
Σ
Σ
Σ
Σ
Σ
List as described in the next
paragraph (Only the measured values of the harmonics are saved.)
Address
Numerical Data
0000
Measured value Total
0004
Measured value DC component
0008
Fundamental component (1st order)
000C
2nd order harmonic component
↓
↓
07D4
500th order harmonic component
Note
Even if the upper limit of the harmonic order under analysis is less than 500th order, data up
to 500th order are saved. The data of the harmonic orders that are not analyzed are saved
as NAN (0x7FC00000).
Appendix 6 Float File Format