Example (General Call)
This example displays the data sequence at the byte level (hexadecimal notation) and indicates the trigger
position. The following notations are used in the figure.
S: Start condition, P: Stop condition, Shading: Compared pattern
• Trigger only on the general call address
Mode
General Call
Second Byte
Not selected
S
27 AE 57 27 FE 98 99
27
P
00
2. Triggers here
A R/W bit
1. Determines whether the first byte is a general call address
• Trigger on a pattern whose second byte address is 06
Mode
General Call
Second Byte
0000 0110
S
06 AE 57 27 FE 98 99
27
P
00
3. Triggers here
A R/W bit
1. Determines whether the first byte is a general call address
2. Determines whether the address pattern is 06
• Trigger when the second byte address matches the specified pattern
Mode
General Call
Second Byte
Master Address, Address pattern: 1010 1011 (0xAB)
Data
Condition: True, Size: 2 bytes, Data pattern: 2
<When the position is not specified>
S
27
AB
AE 57 27 AE 98 99
P
00
4. Triggers here
A R/W bit
1. Determines whether the first byte is a general call address
2. Matches the specified address pattern (AB)
3. Matches the specified data pattern (27 and AE)
<When the position is set to 3>
S
27
AB
AE 57 27 AE 98 99
P
00
5. Triggers here
A R/W bit
1. Determines whether the first byte is a general call address
3. Skips 3 bytes
2. Matches the specified address pattern (AB)
4. Matches the specified data
pattern (27 and AE)