21-1
IM DLM3054-01EN
21 Ethernet Communication (Network)
You can configure TCP/IP parameters and use the Ethernet interface to perform the following tasks.
To use this feature, set the communication interface to Network (from the UTILITY menu, select Remote
Control > Device > Network).
TCP/IP
TCP/IP settings for connecting to an Ethernet network.
Set the IP address, subnet mask, and default gateway.
FTP Server (FTP Server)
You can connect the instrument as an FTP server to a network.
You can connect to the instrument from a PC on the same network and retrieve waveform data.
Mail (Mail)
The action-on-trigger or GO/NO-GO action can be set to mail transmission.
Network Drive (Net Drive)
You can save waveform data and setup data to a network drive.
Network Printer (Net Print)
You can specify a network printer for printing screen captures.
SNTP
The instrument clock can be set using SNTP. The instrument can be configured to automatically adjust its clock
when it is turned on.
To connect a PC to the instrument, use a hub or router, and connect to a network. Do not connect a PC
directly to the instrument.