4-20
IM DLM3054-01EN
Window OR Trigger [ENHANCED]
The instrument triggers on the OR logic of the ranges (window) of multiple trigger sources.
Example
When CH1 = and CH2 =
CH1
Trigger
Upper Level CH1
Lower Level CH1
CH2
Upper Level CH2
Lower Level CH2
Trigger
Trigger
Trigger
Required Condition (Pattern)
Set the trigger sources and trigger slopes.
Trigger Sources
Select the trigger sources from the following: CH1 to CH4 or All. (The selectable settings vary depending on the
model.) If you select All, CH1 to CH4 become trigger sources.
Polarity
Select the polarity from the following.
When the trigger source level enters the specified voltage span (window)
When the trigger source level goes outside of the specified voltage span (window)
Trigger Levels (Levels)
Set the range, HF rejection, noise rejection, trigger position, trigger delay, and trigger hold-off.
Range (Upper Level/Lower Level)
Set the upper and lower levels of the range.
HF Rejection (HF Rejection), Noise Rejection (Noise Rejection)
Set these items for the trigger source.
These items are the same as those of the edge trigger.
Trigger Position (POSITION), Trigger Delay (DELAY), Trigger Hold-Off (Holdoff)
These items are the same as those of the edge trigger.