4-78
IM DLM3054-01EN
A to B(N) Trigger
After the trigger A conditions are met, the instrument triggers when the trigger B conditions are met N times.
CH1
CH2
B(1) B(2) B(3) [trigger]
Condition A met
Condition A: Edge trigger, CH1 = ,
Condition B: Edge trigger, CH2 = , N = 3
Trigger A (A Trigger)
The trigger A conditions are set using the menu that corresponds to the EDGE or ENHANCED key, whichever is
illuminated.
Trigger B (B Trigger)
Configure trigger B (condition B). You can use the following triggers.
Edge trigger, Edge OR trigger, Pattern trigger, Window trigger, Window OR trigger, Serial Bus trigger
• If you set trigger condition A to serial bus trigger, you cannot set trigger condition B to serial bus trigger.
• The trigger level is shared between trigger A and trigger B. If the trigger type of one of the triggers has two
levels (upper level and lower level) and that of the other has one level, the center value of the two levels
and the single level are shared.
Configuring the Trigger B Conditions
The items that you can set vary depending on the A and B trigger types that you selected.
• A Delay B Trigger
Configure these items for trigger B: trigger type, trigger source, trigger level, delay, and other settings.
The trigger type, trigger source, and trigger level settings are the same as those in the EDGE or ENHANCED
key menu.
Delay (Delay)
Selectable range: 10 ns to 10 s
Resolution: 2 ns
• A to B(N) Trigger
Configure these items for trigger B: trigger type, trigger source, trigger level, the number of times condition B
must be met, and other settings. The trigger type, trigger source, and trigger level settings are the same as
those in the EDGE or ENHANCED key menu.
Number of Times Condition B Must Be Met (N Count)
Selectable range: 1 to 10
9
4 Triggering