4-49
IM DLM3054-01EN
Error
The instrument triggers when it detects various types of errors.
• Error Type (Error Type Or)
Select the types of errors to detect from the following. The instrument triggers if any of the selected errors is
detected.
Parity
When a parity error is detected
CRC
When a CRC error is detected
Data Length When the DLC value and the number of data pieces in the data field (data length) do not match
Framing
When the logical value of the stop bit of a field or the stop bit of data is 0
IBS
When the number of IBS bits is 10 or more
Clock
When the clock width exceeds the specified tolerance
PTYPE
The instrument triggers on PTYPE.
Every PTYPE: The instrument triggers on every PTYPE.
No Response: The instrument triggers when a PID is not detected after PTYPE.
ID/Data
The instrument triggers on the AND of SOF, ID, frame information, and data.
• If you specify X in the ID bit pattern, frame information, or data pattern, the condition is assumed to be met
regardless of the corresponding bit or item status.
• In a bit pattern or data pattern setting, if a binary pattern contains any Xs, the corresponding hexadecimal
display will be “$.”
• SOF
The instrument triggers on the start of a CXPI bus signal frame when the trigger condition is set to SOF only.
SOF of ID/Data is always selected as a trigger condition.
• ID
with PTYPE
The instrument triggers on whether a frame contains PTYPE.
• Yes: Triggers on the ID of a frame containing PTYPE
• No: Triggers on the ID of a frame not containing PTYPE
• X: Triggers on the ID regardless of the presence of PTYPE
Bit Pattern (Hex/Bin)
Set the 7-bit ID bit pattern in hexadecimal (Hex) or binary (Bin) notation. The instrument triggers when the
specified bit pattern matches the input signal’s ID bit.
• Frame Information
The instrument triggers on the wakeup and sleep indicators of frame information and CT bit value.
• Wakeup: Set the wakeup indicator bit value to 0, 1, or X.
• Sleep: Set the sleep indicator bit value to 0, 1, or X.
• CT: Set the counter bit value to a value from 0 to 3 or X.
4 Triggering