4-60
IM DLM3054-01EN
Format (Format)
Set the version and signal properties of the SENT signal.
Version
Set the version of the SENT signal.
APR2016: SENT signal complying with the version released in April 2016
JAN2010: SENT signal complying with the version released in January 2010
FEB2008 and older: SENT signal complying with the version released in February 2008 and earlier
Clock Tick
Set the reference clock period of SENT signals. The time between consecutive falling edges of the signal is
counted using this period.
Selectable range: 1.00 μs to 100.00 μs
Resolution: 0.01 μs
Clock Tolerance
Set the clock tick tolerance.
Selectable range: ±1.0% to ±30.0%
Resolution: 0.1%
Data Nibbles
Set the number of data nibbles of fast channel messages.
Selectable range: 1 to 6
Pause Pulse
Select whether to include pause pulses in Fast CH messages. Select this for version APR2016 and JAN2010.
This is fixed to off for version FEB2008 and older.
ON: Pause pulses are included.
OFF: Pause pulses are not included.
CRC Type
Select the CRC type. Select this for version APR2016 and JAN2010. This is fixed to Legacy for version FEB2008
and older.
Recommended: CRC is added using the type recommended in version APR2016 or JAN2010.
Legacy: CRC is added using the type recommended in version FEB2008 and older.
Customize Error Factor
Select the method of detecting
Successive CAL Pulses and Status and Communication.
• Successive Calibration Pulses
Select the method of detecting error type Successive CAL Pulses.
OFF: Errors are not detected.
Preferred Option: The calibration pulse of current frame is compared to the calibration pulse of the preceding
frame to detect errors.
Option 2: The calibration pulse of current frame is compared to the calibration pulse of the last valid preceding
frame to detect errors. However, if three consecutive errors are detected, the third calibration pulse
is considered as valid.
4 Triggering