<About CQ1 autofocus>
CQ1 can perform 2 types of autofocus, which are laser based and
image based.
Laser Autofocus detects surface of bottom plate or cover glass by laser
and set the surface as Z base position. Please note that surface to
detect (Z base positon) is different by the kind of sample vessel. (Refer
to following table.)
Image Autofocus acquires image by designated channel and sets the Z
position that maximum intensity is obtained as Z base position.
Microplate
①
Laser Autofocus detecting surface
(slide glass)
②
Laser Autofocus detecting surface
(microplate, dish, cover glass chamber)
※
③
Image Autofocus detecting position
Dish
Slide
Cover glass chamber
※
In case that laser autofocus
doesn’t operete normally when using microplate, dish
or cover glass chamber, surface to detect would be changed to lower surface of
bottom plate of sample vessel. For details, please refer to page
In
“Map” window, select “Rec Mode” (refer to page
) and select
measuring Area by mouse drug.
Culture medium and others
Bottom plate,
Cover glass
①
②
③
Select “Rec mode”
Drag the Area for
measuring.
Summary of Contents for CQ1
Page 2: ...This document corresponds to the following versions of software Measurement software R1 04 ...
Page 8: ...vi IM 80J01A01 01E Right Side of Utility Box Aperture label ...
Page 24: ......
Page 34: ......
Page 48: ......
Page 177: ... 5 Functions of CQ1 Software 5 81 IM 80J01A01 01E Drag tab to move the window ...
Page 236: ......
Page 238: ......
Page 252: ...7 14 7 Trouble Shooting IM 80J01A01 01E Set sub cover and tighten 2 screws ...
Page 256: ......
Page 266: ......