2-4
<2. Precautions for Use/Installation>
IM 80J01A01-01E
※
CQ1 does not include the pedestal and desk.
Location where enough space is available
If the ventilation openings are blocked, the equipment may
become hot and eventually damaged. Provide enough space
around CQ1 Main Unit and Utility Box.
・
CQ1 Main Unit
Minimum Clearance
Front: 120 mm, Back: 120 mm, Above: 200 mm, Side: 120 mm
Recommended Table Size:
W 700 mm x D 500 mm
x H 700 mm
or larger
Withstand load:
100 kg or larger
・
Utility Box
Minimum Clearance
Back: 50 mm, Above: 50 mm, Side: 50 mm
(Avoid placing on floor directly)
・
Gas Mixer
Minimum Clearance
Back: Provide enough space not to buckle tubes and cables
Side: 50 mm (Avoid placing on floor directly)
Avoid installing at highly humid and / or hot place such as close to
heaters or under direct sunshine, or frequent temperature
changes.
※
1
Location of minimal mechanical vibration
Never set any vibrating devices, such as centrifuge or mixer, on
the same table with CQ1. Doing so may disrupt optical
performance.