<5. Functions of CQ1 Software>
5-55
IM 80J01A01-01E
(3) XY Stage
Set moving speed of XY stage. [Edit] button is to select XY stage
speed.
①
Speed
Select XY stage speed. Default setting is the fastest. If samples
are easy to ablate from sample vessel, slow setting is
recommended.
To apply, click
③
Set
②
Cancel
Cancel XY stage speed change
③
Set
Set XY stage speed
(3) Laser-Power Calibration
Click
“Calibrate” button to show “Laser-Power Calibration” dialog, and
click
“Yes” to proceed calibration, but no calibration by clicking “No”.
Laser power calibration is performed to maintain excitation laser power
irradiated to samples against
“Excitation Power” set in “Imaging”
window (refer to
). It takes a few minutes per 1
laser line to calibrate.
In case that error occurs in laser power calibration, optical adjustment
is needed. Please contact the dealer.