Frequency controller
Switching frequency control is used, e.g. for controlling dosing pumps.
In contrast to the pulse-width controller, the frequency controller is not
modulated by the pulse width, but by the switching frequency of the out-
put signal. Depending on the position of the measured value in the pro-
portional range, the relay is switched more often or less often.
Fig. 5-8
Relay output of the frequency controller
While the selected switching duration (t
on
= 0.3 s) always remains con-
stant, the switching frequency at which the relay switches changes de-
pending on the measured value.
If the measured value is at the end of the proportional range (
End
value
), the switching frequency is higher.
If the measured value is at the beginning of the proportional range
(
Start value
), the switching frequency is low.
Time [s]
t
= 0.3 s
On
Relay
Off
On
T