TresCon UNO
Installation
3 - 15
ba43134e08
05/2017
Switching voltages and switching currents on the relay contacts
must not exceed the values specified in chapter 7 T
.
Protect electrical circuits against currents that are too high with an
electrical fuse.
The mounting frame may only be opened after the installation if all
external voltages have been previously switched off.
The analyzer has programmable control, indicator and alarm contacts.
Warning
If the load at an indicator contact is inductive, an RCD protective circuit
is required! For an example of the circuit, see Fig. 3-14.