TresCon UNO
Operation
5 - 97
ba43134e08
05/2017
The permeate switch is executed in the following steps:
Note
You can adjust the permeate state to the state of a relay (relay configu-
ration section 5.3.8 R
on page 5-47).
Permeate switch
measured value display
Step
Instrument status/action
State
Duration
in min
1
Permeate 1 is fed to the analy-
zer module
MeasureSample 1
2
Start measurement of
permeate 1
MeasureSample 1 10 ... 999
3
Stop measurement of
permeate 1
4
Permeate 2 is fed to the analy-
zer module
MeasureSample 2
5
Start measurement of
permeate 2
MeasureSample 2 10 ... 999
5
Stop measurement of
permeate 2
Permeate state P1/P2
Relay state (optional)