![wtw TresCon UNO A111 Operating Manual Download Page 198](http://html1.mh-extra.com/html/wtw/trescon-uno-a111/trescon-uno-a111_operating-manual_1000694198.webp)
Appendix
TresCon UNO
8 - 20
ba43134e08
05/2017
8.4.3
Commands for PROFIBUS
PROFIBUS connection
PROFIBUS record (PBP)
Note
Further information on wiring, configuration and the record can be
found in the "PKV 30-DPS instrument manual" and the manual "Con-
nection of WTW measured value recording systems to the PROFIBUS-
DP" that are enclosed in the PKV 30-DPS converter.
PKV 30-DPS
RS 485
Profibus DP
PC
WTW monitor
with RS 485 option
TresCon
Handshake
(binary)
this byte must be toggled (1<->0),
in order to trigger a data transfer on the
RS485 side
Slave address
(binary)
RS485 address of the measuring transmitter
(1...31)
Length information
(binary)
Length of the ASCII message in bytes (1...61)
ASCII message
The return values in the response are separa-
ted by spaces.
The length of the response is variable.
PBP overhead
PBP message
64 bytes
(fixed length)
PBP overhead
1 byte
1 byte
1 byte
n bytes
Handshake Slave address
RS485
Length information
n ASCII
ASCII
message
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......