Appendix
TresCon UNO
8 - 4
ba43134e08
05/2017
Wiring of the
BUS
(Slave operation)
Up to 31 TresCon analyzers can operate together on the BUS.
The last instrument on the BUS must provide the terminating resistor
(see Fig. 8-1). If only one instrument is used, this is also the last instru-
ment and must provide the terminating resistor.
Fig. 8-1
TresCon in the BUS
This setting can be made in the software of the last TresCon in the
General Parameters
menu (bus termination).
Note
The parameterization of the interface is described in section
5.5.2 RS485 (S
, on page 5-62.
Terminating resistor