Maintenance
TresCon UNO
6 - 14
ba43134e08
05/2017
Defective key
If a key is defective, the TresCon UNO
Key Test
remains in the service
mode. Proceed as follows to return to operation again.
Attention
If safe operation is no longer possible due to the defective key, remove
the TresCon UNO from operation and protect it against being returned
to service!
5
Leave the menu with
4
or
3
.
1
Switch off the TresCon UNO at the power switch (on the moun-
ting element).
2
Switch on the TresCon UNO again at the power switch (on the
mounting element) after a brief waiting period. The status of
the modules (measuring, AutoCal, AutoClean) before the inst-
rument was switched off is restored.
3
You should have the key field repaired as quickly as possible
by a specialist authorized by WTW or contact the service
department of WTW directly!