TresCon UNO
Installation
3 - 7
ba43134e08
05/2017
Connecting the overflow
vessel
Fig. 3-8
Tubes of the overflow vessel
Note
The back connection piece of the overflow vessel is for aeration. It must
not be closed.
13
Insert both free ends of the drain set in a suitable, pressure-free
drain.
14
Connect the overflow vessel to the following connectors (see
Fig. 3-8):
– Connect the sample or permeate tube (black tube, inner dia-
meter = 6 mm) to the front connection piece of the overflow
vessel.
– Connect the sample overflow (center connection piece) to
the drain set.
Tube: Inner diameter = 6 mm, black.
– Lead the permeate tube of the analyzer module (black
tubing, inner diameter = 1.5 mm) through the left-hand dril-
ling in the TresCon UNO enclosure bottom and connect it to
the fitting of the overflow vessel.
drain
O
I
P1
to measuring
module
sample
(permeate)
sample
(permeate)