![wtw TresCon UNO A111 Operating Manual Download Page 183](http://html1.mh-extra.com/html/wtw/trescon-uno-a111/trescon-uno-a111_operating-manual_1000694183.webp)
TresCon UNO
Appendix
8 - 5
ba43134e08
05/2017
8.3.2
Operating modes
General information
Basically, there are 2 different operating modes:
Master operation: Output of measured values
Slave operation: Bidirectional operation via the BUS
Application cases
Each operating mode is designed for a specific application case:
The interface is always active in the Master operating mode. The data
is output to one data terminal (terminal or printer); the BUS connection
(up to 31 devices) is possible in Slave operation only.
The Slave operating mode is offered exclusively with the BUS, even if
only a single instrument is present.
8.3.3
Master operation
The Master operation is for applications that have a terminal (program)
or printer, i.e. data is sent continually from the instrument (Master)
during the measuring mode.
8.3.4
Slave operation
The measuring instruments ("Slave") operate together with a control
computer ("Master") according to the Master/Slave principle in this ope-
rating mode. The modules are addressed by the control computer
under their address and execute the commands that are handed over.
To do so, the control computer must also take over the control of the
level converter (send/receive DTR/DSR).
The communication procedure is divided into the command section and
the response section.
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......