TresCon UNO
Installation
3 - 33
ba43134e08
05/2017
Fig. 3-28 Connecting the permeate feed and outlet tubes
Attention
The drain must be pressure free!
Note
The back connection piece of the overflow vessel is for aeration. It must
not be closed.
3
Connect each of the two overflow vessels to the following con-
nectors (see Fig. 3-28):
– Connect the sample or permeate tubes (black tube, inner
diameter = 6 mm) to the front connection piece of the over-
flow vessel.
– Connect the sample overflow (center connection piece) to
the drain set.
Tube: Inner diameter = 6 mm, black.
4
Lead the permeate tube of the analyzer module (black tube,
inner diameter = 1.5 mm) through the left-hand drilling in the
bottom of the TresCon UNO enclosure and connect it to the
center fitting of the three/two way valve on the permeate
switch.
to measuring
module
O
I
Pumpschlauch Standard/Probe
blau / blau / blau
Pumpschlauch Reagenz
orange / gelb / orange
Pump Tube Standard/Sample
blue / blue / blue
Pump Tube Reagent
orange / yellow / orange
TresCon PO4-P
Reinigungslösung
Cleaning Solution
H O dest.
H O dest.
2
2
Standard B
Standard B
Reagenz
Reagent
Ablauf
blau
blue
V3
V2
V4
Thermoblock
Probe
Sample
V1
gelb
yellow
orange
orange
grün
green
blau
blue
rot
red
grün
green
P1
P2
P3
sample
(permeate)
sample 2
sample 1
drain
Probe 1
Probe 2
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......