TresCon UNO
Appendix
8 - 19
ba43134e08
05/2017
msalarmcall
This command defines for different components that a call should be
made if a concentration value is exceeded.
Syntax:
MsalarmcallNumber,Interval,Value1 <CR>
Sample transmission:
The callback is deactivated by sending the command,
"msalarmcall0,0,0"
. It can also be deactivated in the menu for ini-
tializing the modem (Deactivating of modem). When doing so, all set-
tings are deactivated.
Number
Modem number of the receiving address to be called
Interval
Interval in minutes for the repetition of the call if the
value is exceeded permanently
Value1
Concentration value (call in the case of exceeding)
------------------------------------------------------------
05.08.03 13:22:08
Alarm call
TresCon-Modem
Module 1: 10.55 > 10.00 mg/l NOx
+++
NO CARRIER