![wtw TresCon UNO A111 Operating Manual Download Page 118](http://html1.mh-extra.com/html/wtw/trescon-uno-a111/trescon-uno-a111_operating-manual_1000694118.webp)
Operation
TresCon UNO
5 - 62
ba43134e08
05/2017
5.5.2
RS485 (Slave) interface parameters
The RS485 interface can be operated in either the "Slave" or "Master"
mode.
Note
The chapter 8 A
contains general information on the applica-
tions of the RS485 interface and provides examples of these applica-
tions.
Data
The settings when the instrument is delivered are marked in bold:
Viewing/changing
parameters
Parameter
Values
Mode
Slave,
Master
Baud rate
1200, 2400, 4800,
9600
, 19200, 38400
Data bits
7,
8
Stop bit
1
, 2
Parity
None
, Even, Odd
Bus address
0,
1
, 2 ... 31
Bus termination
Yes
, No
1
Measuring mode
Call up the
General Parameters
menu: Simultaneously press
the keys,
4
and
2
.
2
Select the
Interfaces
menu item with
68
and confirm
with
3
. The prompt
Presentation of data/Change
appears
.
3
Select the
Change
mode with
68
and confirm with
3
.
4
Input the PIN digit-by-digit with
68
. Change the digit
with
57
and confirm with
3
. The
Interface Parame-
ters
menu appears in the display providing a view of the RS232
interface.
5
Select the selection mode with
3
and select the RS485
interface with
68
.
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......