RS485 Master bus protocol
String in transmission buffer
(max. 61 chars./block)
Enter length
Calculate and enter CRC
Initialize input buffer
Send message
Start timeout B
Wait for characters to be
received
Yes
No
Repeat-Counter = 1
Repeat-Counter
+1
Repeat-Counter > 3
IF (timeout)
IF (timeout)
Load length byte into buffer
Load address byte
Start timeout A
Wait for characters to be
received
length > 0
Start timeout A
Wait for characters to be
received
IF (timeout)
Load byte into buffer
Length = length - 1
1/2 timeout B
Wait
Yes
Response invalid
Error status:
No response
Yes
Yes
CRC
CRC correct?
Address
--
Slave address
Continuation bit set?
Response is valid
Error status = Bit 5
Slave address off
END
For write commands, the
response is an empty string.
The error status within the
Slave address is valid!
No
No
Yes
Yes
No
No
No
Timeout times:
– A = 3 byte transmis-
sion times (approx. 3.1
ms with 9600 bauds)
– B= approx. 1s
Summary of Contents for TresCon UNO A111
Page 6: ...Inhaltsverzeichnis TresCon UNO 0 4 ba43134e08 05 2017 9 1 Abbreviations 9 1 9 2 Index 9 1...
Page 18: ...Safety TresCon UNO 2 4 ba43134e08 05 2017...
Page 56: ...Initial commissioning TresCon UNO 4 4 ba43134e08 05 2017...
Page 133: ...TresCon UNO Operation 5 77 ba43134e08 05 2017 9 Select Yes or No with 68 and confirm with 3...
Page 167: ...TresCon UNO Maintenance 6 11 ba43134e08 05 2017 6 Leave the menu with 4...
Page 202: ...Appendix TresCon UNO 8 24 ba43134e08 05 2017...
Page 207: ......