NOx-N
Technical data
5 - 1
ba43110e05
01/2011
5
Technical data
Note
All specifications of the measuring accuracy refer to the use of suitable
standard solutions.
Note
The TresCon NOx-N continues to measure outside the measuring
range, although with lower accuracy. A note appears on the display.
Measuring principle
4-wavelength photometric process
Measuring method
Individual absorption of nitrate/nitrite
Measuring ranges
mg/l
µmol/l
NOx-N
0.1 - 60.0
0 - 4000
NOx
0.1 - 250
0 - 4000
Resolution
Resolution
Range
mg/l NOx-N
0.1 mg/l
0.1 ... 60.0 mg/l
Resolution
Resolution
Range
mg/l NOx
0.1 mg/l
0.1 ... 99.9 mg/l
1 mg/l
100 ... 250 mg/l
Resolution
Resolution
Range
µmol/l
NOx-N / NOx
1 µmol/l
0 ... 4000 µmol/l
Measuring
accuracy
Absolute error ± 2 % of measured value
± 0.4 mg/l NOx-N
Process variation coefficient 2 %
Response time
Approx. 30 seconds
(after a change in concentration at the input)
Measuring interval
Selectable: Continuous, 5 min, 10 min,
15 min, 20 min, 25 min, 30 min
Type of calibration
Automatic zero-point balancing
Fixed factory calibration
Calibration interval
Selectable: 1, 2, 3, 6, 12 and 24 hours
Reference time
Selectable in 30 min intervals
Measurement
Calibration
(AutoCal)