Appendix
NOx-N
6 - 4
ba43110e05
01/2011
TXRCALI
The data of the last zero point calibration (AutoCal) is read.
Instrument response:
If the last calibration was erroneous, the data of the erroneous
calibration and the last valid calibration is displayed. The data of the
erroneous calibration is marked with an "err" on the right-hand side of
the time.
TXRPOLYNOM
The data of the last multi-point calibration is read.
Instrument response:
If the last calibration was erroneous, the data of the erroneous
calibration and the last valid calibration is displayed. The data of the
erroneous calibration is marked with an "err" on the right-hand side of
the time.
-----------------------------------------------------------
Calibration data
NOx-N Module 1
22.04.03 15:26:24
-----------------------------------------------------------
Calibration data of 22.04.03 15:25:56
E1H2O: 1.66 mE
E2H2O: -74.33 mE
-----------------------------------------------------------
TresCon>
-----------------------------------------------------------
Calibration data
NOx-N Module 1
22.04.03 15:28:56
-----------------------------------------------------------
Calibration data of 05.07.02 11:08:36
Standard A : 10.0 mg/l NO3
Standard B : 50.0 mg/l NO3
Standard C : 150.0 mg/l NO3
Standard D : 250.0 mg/l NO3
E1A: 232.50 mE E2A: 1.85 mE
E1B: 907.80 mE E2B: 5.74 mE
E1C: 1930.30 mE E2C: 14.42 mE
E1D: 2446.80 mE E2D: 22.20 mE
a0: -0.00002
a1: 36.83043
a2: 30.13768
a3: -16.74767
a4: 6.27163
g0: 0.00081
g1: 0.00009
h: 1.200
-----------------------------------------------------------
TresCon>