Operation
NOx-N
2 - 16
ba43110e05
01/2011
2.8
Displaying/changing the display parameter
You can display or change the dimension/citation form in this menu.
Converting the citation
forms and units
The reference value is NOx [mg/l]. Using this value, the displays are
calculated in another unit or cited form by multiplying them with the
specified factor:
Effects of the
conversion
The following table demonstrates what effects the adjusted citation
form has where:
Dimension
Citation forms
NOx
NOx-N
mg/l
1.0
0.2259
µmol/l
16.13
16.13
Measured value
Yes
Measuring range
Yes
Input and display of Cnominal
Yes
Relay functions
Yes (
General
Parameters
menu)
Recorder functions
Yes (
General
Parameters
menu)
Parameter
Values
Citation form
NOx-N
NOx
Dimension
mg/l
µmol/l
1
Measuring mode
Select TresCon NO
X
-N with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Presentation of data
menu item with
68
and
confirm with
3
. The prompt
Presentation of data/Change
appears.
Inputs