Maintenance
NOx-N
3 - 14
ba43110e05
01/2011
3.6
Displacing/replacing pump tubings
The pump tubings must be displaced or replaced at regular intervals.
The TresCon has timers for this that indicate the maintenance activities
required when the relevant service intervals expire. This service menu
"guides" you through the necessary steps. The service routine consists
of two parts to enable the work on the instrument to be performed at the
correct time.
Note
The instrument indicates when the pump tubings must be displaced/
replaced by a timely message (service intervals).
1
Measuring mode
Select the NOx-N measuring module with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Displace/replace pump tubings
menu item with
68
and confirm with
3
. The time until the necessary
displacing/replacement based on the service intervals appears
on the display.
5
Return to the service menu with
4
.
Displaying service
intervals