NOx-N
Operation
2 - 27
ba43110e05
01/2011
2.13
Dilution parameter
You can select the dilution of the sample in this menu.
How it functions
If the sample is diluted before it reaches the TresCon, the dilution factor
can be selected. The actual concentration value is displayed.
Example
The dilution is set to 1+3 (1 part test sample, with the aid of
PurCon diluted with 3 parts H2O).
The analyzer measured 1.00 mg/l NOx-N and multiplies the value with
the factor 4. 4.00 mg/l NOx-N appears on the display.
Note
The measuring range corresponding to the resolution is given in the
parameter ticket. By using a dilution device before the TresCon, the
dilution error becomes part of the total error.
Parameter
Values
Dilution
no, 1+1, 1+2, 1+3, 1+4, 1+5, 1+6, 1+7, 1+8, 1+9,
1+11, 1+13, 1+15, 1+17, 1+19
1
Measuring mode
Select TresCon NO
X
-N with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Dilution
menu item with
68
and confirm with
3
. The prompt
Presentation of data/Change
appears.
4
Select the
Change
mode with
68
and confirm with
3
.
5
Enter the PIN digit-by-digit with
68
. Change to the next
digit with
57
and confirm with
3
.
The analyzer switches into the stop mode.
Changing the dilution
parameter