Maintenance
3 - 11
ba43110e05
01/2011
3.5
Emptying the system
This service routine empties the wet chemical system of the instrument.
z
Dist. H2O
z
Cleaning solution
Cleaning solution causes severe corrosion
Do not allow this substance to get into the hands of children. Wear
suitable protective gloves and goggles/face mask when working with
this substance.
If this substance gets into the eyes, rinse thoroughly with water and
consult a doctor.
Warning
Only trained personnel is allowed to work on the analyzing section.
Always follow the warning instructions on the chemical containers and
the safety datasheets!
Always use protective equipment!
Note
If the TresCon is taken out of service for more than 24 hours, the
Empty
system
service routine must be performed!
Duration
Approx. 1 ½ minutes
Caution
The pump empties the system via the tubes. If the canisters are not
removed, the cleaning solution and dist. H2O will become unusable!
You cannot quit the procedure.
Step
Instrument status/action
1
Flush back dist. H2O
2
Flushing back cleaning solution
3
Flushing back sample
Wear protective
goggles
Wear protective
gloves
Wear protective
clothing
Program sequence