Maintenance
NOx-N
3 - 10
ba43110e05
01/2011
Caution
Only use distilled water without any contamination (softener or similar)!
Dist. H2O is also used for zero point calibration. Contaminated dist.
H2O causes incorrect measurement results.
5
Place a filled container of dist. H2O on the tray.
6
Connect the container. Follow the color coding (red)!
7
Confirm the prompts each time with
3
after checking. After
the last confirmation, the procedure begins immediately and
the remaining duration appears on the display.
8
At the end of the procedure,
Service finished
appears on the
display.
9
Return to the service menu with
3
.