Operation
NOx-N
2 - 14
ba43110e05
01/2011
Caution
Only change Cnominal in special cases!
Read the prerequisites in this section for more information!
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Manual correction
menu item with
68
and
confirm with
3
. The prompt
Presentation of data/Change
appears.
4
Select the
Change
mode with
68
and confirm with
3
.
5
Enter the PIN digit-by-digit with
68
. Change to the next
digit with
57
and confirm with
3
. The analyzer
switches into STOP mode.
6
The
Manual correction
parameter is selected. Switch into input
mode with
3
.
7
Select
On
with
68
and confirm with
3
. The cursor
skips to the line for the input of Cnominal.