Maintenance
NOx-N
3 - 36
ba43110e05
01/2011
Preparing the
calibration
To calibrate, you need the following four calibration standards (for
ordering information, see section 6.1 A
Cleaning solution causes severe corrosion
Do not allow this substance to get into the hands of children. Wear
suitable protective gloves and goggles/face mask when working with
this substance.
If this substance gets into the eyes, rinse thoroughly with water and
consult a doctor.
Starting calibration
Ca
10 mg/l NO3
-
Cb
50 mg/l NO3
-
Cc
150 mg/l NO3
-
Cd
250 mg/l NO3
-
1
Check that the dist. H2O is connected and, if necessary,
connect it. Only use distilled water free of contamination (e.g.
softeners).
2
Pull the sample tubing out of the overflow vessel.
3
Insert the open end of the tube into a collecting container.
1
Measuring mode
Select the NOx-N measuring module with
68
.
2
Call up the
Parameters NO
X
-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Calibration
menu item with
68
and confirm
with
3
.
5
Input the PIN digit-by-digit with
68
. Change the digit
with
57
and confirm with
3
. A checklist appears on
the display.