NOx-N
Maintenance
3 - 21
ba43110e05
01/2011
3.6.2
Displacing the pump tubings
For this service, the consumables are first pumped back out of the
system via the sample tube.
Caution
Only trained personnel is allowed to work on the analyzing section.
Caution
After step 5, you can no longer quit the procedure!
Displacing the pump
tubings
Reset the
Displace pump tubings
timer
1
Measuring mode
Select the NOx-N measuring module with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Displace/replace pump tubings
menu item with
68
and confirm with
3
.
5
Select the
Displace pump tubings
service with
68
and
confirm with
3
. The 1st part of the checklist appears on the
display.
Effects
Displacing the
pump tubings