NOx-N
Maintenance
3 - 9
ba43110e05
01/2011
3.4
Replenishing dist. H2O
This routine replenishes the wet chemical system with dist. H2O
.
The
distilled water must be free of contamination (softener orsimilar) as it is
also used for the zero-point calibration.
Duration
Approx. ½ minute
Note
You cannot quit the procedure.
9
Return to the service menu with
3
.
Step
Instrument status/action
1
Rinsing dist. H2O
Program sequence
1
Measuring mode
Select the NOx-N measuring module with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Replenish H2O dest
menu item with
68
and
confirm with
3
. A checklist appears on the display.
Starting the procedure