NOx-N
Maintenance
3 - 7
ba43110e05
01/2011
3.3
Replenishing the cleaning solution
This routine fills the wet chemical system with cleaning solution.
Cleaning solution causes severe corrosion
Do not allow this substance to get into the hands of children. Wear
suitable protective gloves and goggles/face mask when working with
this substance.
If this substance gets into the eyes, rinse thoroughly with water and
consult a doctor.
Warning
Only trained personnel is allowed to work on the analyzing section.
Always follow the warning instructions on the chemical containers and
the safety datasheets!
Always use protective equipment!
Duration
Approx. ½ minute
Note
You cannot quit the procedure.
Step
Instrument status/action
1
Rinsing cleaning solution
Wear protective
goggles
Wear protective
gloves
Wear protective
clothing
Program sequence
1
Measuring mode
Select the NOx-N measuring module with
68
.
2
Call up the
Parameters NOx-N
menu with
3
. The main
menu appears on the display.
3
Select the
Service
menu item with
68
and confirm with
3
. The service menu appears.
4
Select the
Replenishing cleaning solution
menu item with
68
and confirm with
3
. A checklist appears on the
display.
Starting the procedure