What to do if...
NOx-N
4 - 4
ba43110e05
01/2011
4.2
Messages
A text appears on the display although the measuring mode is not,
however, quit. Example:
Replenish dist. H2O
. You can perform the
relevant action later.
Note
The messages disappear when the appropriate measure has been
performed (see chapter 3 M
Cause Action
An error occurred when
writing parameter data.
Confirm the prompt with
3
; contact the
WTW service department in order to
measure.
Cause Action
Fluid level in the
container has fallen
below 4 cm.
Replace the container with a full one.
Start the
Replenish dist. H2O
service.
Cause Action
Fluid level in the
container has fallen
below 4 cm.
Replace the container with a full one.
Start the
Replenish cleaning solution
service.
Write error: Use factory
setting
Fill dist. H2O
Replenish
cleaning solution